A Comparative Study of Different Methods of Determining Kt/V
Article History
-
Received: ,
Accepted: ,
,
Abstract
Background
Kt/V has been used as a standard for determining dialysis adequacy. Studies comparing Kt/V calculated by ultraviolet (UV) absorbance-based online clearance monitoring (OCM) with other methods are scanty. The study aims to assess agreement between OCM and other methods of Kt/V estimation and to evaluate the impact of gender on estimating Kt/V by different methods.
Materials and Methods
This was a cross-sectional study including 50 patients with end-stage renal disease (ESRD) undergoing maintenance hemodialysis (HD) on a B Braun Dialog HD machine integrated with Adimea technology with a 1.4 m2 dialyzer. Kt/V was calculated by three methods, namely, Daugirdas formula, nomogram method, and OCM by UV absorbance.
Results
The mean age was 56 ± 11.78 years, with 68% being males. The mean Kt/V values estimated by OCM, Daugirdas formula, and nomogram were 1.43, 1.29, and 1.26, respectively. Kt/V measured by OCM (UV absorbance) was significantly higher compared to the other two blood-based methods. However, agreement analysis using Bland-Altman plots demonstrated the presence of systematic bias between OCM Kt/V and Daugirdas Kt/V. The observed differences between OCM Kt/V and other methods likely reflect methodological variations rather than any definitive differences in delivered dialysis dose. Linear regression analysis demonstrated that females and males with lower body weight had significantly higher Kt/V. Although visual inspection of Bland-Altman plots also suggested a possible trend, regression analysis did not demonstrate statistically significant proportional bias.
Conclusion
OCM Kt/V by UV absorbance overestimates Kt/V compared with other methods, with moderate agreement, limiting its clinical interchangeability.
Keywords
Introduction
Monitoring and providing adequate hemodialysis (HD) is an essential aspect of the treatment of patients with end-stage renal disease (ESRD). Conventionally, Kt/V has been used as a standard for determining dialysis adequacy.1 Despite its shortcomings, Kt/V still remains an important parameter to measure if adequate dialysis has been delivered.1,2 There are three methods of calculating Kt/V, which include estimation by the Daugirdas single pool(sp)Kt/V formula, the Nomogram, and the online clearance monitoring (OCM) method.3 The latter can be done by two methods, i.e., ionic dialysance and ultraviolet (UV) absorbance-based method.4 Both these OCM methods offer the advantage of being non-invasive and providing real time assessment in each HD session.5 However the OCM Kt/V by ionic method faces limitations as it involves calculation of total body water (TBW) or urea distribution volume (V) using anthropometric measurements, age and gender which is prone to irregularities especially in patients with ESRD.6,7 OCM by UV absorbance avoids the estimation of V. We thus aimed to compare different methods of calculating adequacy of HD in terms of Kt/V and to assess agreement of OCM Kt/V with other two methods. We also evaluated the impact of gender in estimating Kt/V by different methods.
Materials and Methods
This was a cross-sectional study done at the HD unit of a tertiary care referral hospital over a period of 6 months from March-August 2025. The study was approved by the Institutional Ethics Committee of St. John’s National Academy of Health Sciences, Bangalore, (#222/2024, 21/08/24). Informed consent was taken from all patients. 50 hemodynamically stable patients with ESRD aged >18 years on regular maintenance HD were included in this study. Patients were considered hemodynamically stable if they had stable blood pressure (BP) of systolic BP ≥90 mmHg (or mean arterial pressure ≥65–70 mmHg) that is sustained without intervention, no symptomatic intradialytic hypotension requiring medical intervention, drop in systolic BP >30 mmHg or mean arterial pressure drop of >20 mmHg, no session interruption, and no need for blood-flow or UF adjustment during the measured dialysis session.8 Critically ill patients with sepsis, decompensated heart failure, decompensated liver disease, malignancies, and hemodynamically unstable patients were excluded from the study. The UF rate threshold was not set, and the ultrafiltration (UF) rate per session was determined based on intradialytic weight gain.9 Also, dialysis sessions with a sharp rise in Kt/V decay absorbance curve in the OCM method, due to causes such as altered blood flow, filter clotting, or recirculation, were excluded.
Kt/V was calculated by three methods, namely by Daugirdas single pool(sp)Kt/V formula,10 nomogram,11 and OCM by UV absorbance-based method.12 HD was done on a B Braun Dialog machine integrated by Adimea technology with a 1.4 m2 dialyzer. The duration of dialysis was four hours per session. The blood flow rate was fixed at 250 mL/min, and dialysis flow rate at 500 mL/min for the measured session. If blood flow rate had to be changed due to hemodynamic alterations, then the session was not included for measurement.
For Daugirdas formula and nomogram method, each patient underwent blood sampling twice for calculating the urea reduction ratio (URR). The first sample was taken before the initiation of HD, and the second sample was obtained at the end of the same session after setting the UF rate to zero, reducing blood pump flow to less than 100 mL/min for 2–3 min, and then stopping pump flow.
The nomogram method involved plotting the URR and UF/W (Ultrafiltration by post dialysis weight) on a specially designed nomogram, which gave a value of Kt/V.11 The second generation Daugirdas formula of estimating single pool Kt/V10 (− ln (R-0.008 × t) + (4–3.5 × R) × (UF/W) proposed by J.T. Daugirdas was used for formula method wherein, R= post-dialysis blood urea nitrogen (BUN)/pre-dialysis BUN, t = dialysis time (hours), - ln = negative natural logarithm, UF/W = ratio of ultrafiltration volume in litres to post -dialysis weight in kg. V was calculated by the Watson formula.13
The OCM method used Adimea® technology, which is integrated in the B Braun Dialog HD machine and has been validated for HD.14 The only input required is the patient’s pre-dialysis weight. We determined OCM spKt/V as it is most commonly used. It is equipped with continuous monitoring of effluent dialysate by UV absorbance at 280 nm. It utilizes the principles of spectroscopy, working by transmitting UV light through the spent dialysate. The particles contained in the dialysate, which are removed from plasma, absorb the light depending on the concentration, which is then measured by a sensor. It has been proven that the UV absorption measurements can be used to determine the dialysis dose, as there is a very close linear correlation between the measured UV absorption signal and the urea in the spent dialysate.12,15 The OCM method calculates Kt/V using continuous UV absorbance monitoring of spent dialysate under standardized dialysis conditions.15 Time averaging was done by continuously measuring the UV absorbance of urea in spent dialysate throughout the session. It then calculated Kt/V by fitting into the decay absorbance curve, which reflects urea removal.12,15 To minimize methodological variability, dialysis duration, blood flow rate, dialysate flow rate, and dialyzer surface area were kept constant throughout the study. If the dialysis session was interrupted due to recirculating blood during fistula needle adjustment or the patient going to the bathroom, the HD machine was put into bypass to avoid errors.
Statistical analysis was carried out using the R statistic software system. Mean ± standard deviation was calculated for quantitative data like age (yrs), height (cm), dry weight (kg), pre dialysis BUN (mg/dL), pre dialysis weight (kg), body surface area (m2), UF (L), UF/W, post-dialysis weight (kg), URR (%), urea distribution volume (V in litres), OCM Kt/V by UV absorbance, Daugirdas formula Kt/V and nomogram Kt/V. Frequencies and percentages were calculated for qualitative data like gender. One-way ANOVA was used to compare mean Kt/V values across the three methods, while the Pearson correlation coefficient was used to assess linear association.
Agreement analysis was done using Bland-Altman plots and Concordance Correlation Coefficient (CCC) to determine agreement and systematic bias between OCM Kt/V and nomogram Kt/V with the comparator method of Daugirdas formula Kt/V. Kappa statistics were used for the evaluation of categorical agreement among the adequacy classification. The presence of proportional bias was formally assessed by linear regression of the differences against the mean.
Results
The baseline characteristics and comparison between the characteristics of males and females, including different methods of Kt/V, have been shown in Table 1. The males and females had no significant age differences. However, males had physiologically expected higher height, dry weight, body surface area, and post-dialysis weight. V was significantly higher in males (36.67 ± 4.48 vs. 28.54 ± 3.23; p <0.001), resulting in statistically significant lower Kt/V by all three methods. The mean Kt/V values obtained by OCM (UV absorbance), Daugirdas formula, and nomogram were 1.43, 1.29, and 1.26, respectively [Table 1]. Females in this study demonstrated significantly higher URR% (p = 0.004), OCM Kt/V (p <0.001), Daugirdas formula Kt/V (p = 0.007), and nomogram Kt/V (p = 0.011). Thus, there was a significant difference between OCM Kt/V for male and female patients based on ANOVA [Table 1].
| Parameter | Male (Mean ± SD) | Female (Mean ± SD) |
|---|---|---|
| Age (yrs) | 55.12 ± 8.75 | 57.05 ± 13.08 |
| Height (cm) | 166.63 ± 9.76 | 157.50 ± 11.96 |
| Dry weight (kg) | 64.73 ± 12.50 | 55.99 ± 11.96 |
| Body surface area (m2) | 1.71 ± 0.16 | 1.54 ± 0.16 |
| Pre-dialysis BUN (mg/dl) | 46.96 ± 13.11 | 42.22 ± 9.49 |
| Urea reduction ratio (%) | 61.73 ± 7.78 | 68.60 ± 7.03 |
| Urea distribution volume (V) | 36.67 ± 4.48 | 28.54 ± 3.23 |
| Ultrafiltration volume (L) | 3.79 ± 0.73 | 2.88 ± 1.13 |
| Post-dialysis weight (W) | 65.91 ± 13.06 | 56.28 ± 12.51 |
| UF/W | 0.05 ± 0.01 | 0.04 ± 0.02 |
| OCM Kt/V | 1.33 ± 0.22 | 1.63 ± 0.32 |
| Formula Kt/V | 1.22 ± 0.25 | 1.45 ± 0.30 |
| Nomogram Kt/V | 1.20 ± 0.24 | 1.39 ± 0.22 |
SD: Standard deviation, BUN: Blood urea nitrogen, URR: Urea reduction ratio, UF/W: Ultrafiltration by post dialysis weight, OCM: Online clearance monitoring
The mean OCM-derived Kt/V was 1.43 ± 0.29, whereas the mean formula-derived Kt/V was 1.29 ± 0.28. Similarly, the mean nomogram-derived Kt/V was 1.26 ± 0.25. Comparison of OCM-derived Kt/V with nomogram-derived Kt/V demonstrated a statistically significant difference (1.43 ± 0.29 vs. 1.26 ± 0.25, p = 0.003) . The comparison between OCM Kt/V and nomogram Kt/V with the comparator Daugirdas formula Kt/V is as depicted in linear regression graphs [Figures 1a and b]. OCM Kt/V demonstrated a strong positive correlation with Daugirdas formula Kt/V (Pearson’s r = 0.731). In contrast, the difference between nomogram-derived and formula-derived Kt/V values was not statistically significant (1.26 ± 0.25 vs. 1.29 ± 0.28, p = 0.539). Correlation between all three methods was statistically significant as measured by Pearson correlation and linear regression model. OCM Kt/V for both male and female patients correlated well with Daugirdas formula, as shown in [Figures 2a and b]. Males with lower body weight had higher OCM Kt/V (r = -0.554, p <0.001).


However, as correlation and linear regression can be high even when there is systematic bias, agreement analysis was done. Comparison between nomogram and Daugirdas Kt/V demonstrated excellent agreement (mean bias = 0.033), reinforcing reliability (CCC = 0.949, 95% CI: 0.917–0.969), indicating near-perfect concordance. The 95% limits of agreement (LoA) ranged from -0.126–0.192 [Figure 3]. The observed strength of agreement was very high between Daugirdas and the nomogram Kt/V (Kappa = 0.87, p <0.001) [Table 2].

| Method | Observed agreement (%) | Kappa | p value |
|---|---|---|---|
| OCM vs. Daugirdas formula Kt/V | 76 | 0.45 | <0.001 |
| Nomogram vs. Daugirdas formula Kt/V | 94 | 0.87 | <0.001 |
OCM: Online clearance monitoring
In contrast, OCM Kt/V comparison to Daugirdas Kt/V showed only moderate agreement (mean bias = 0.133) with an observed agreement of 76% with wider and lower confidence interval (CCC = 0.660, 95% CI: 0.490–0.782), indicating variability in consistency between methods. The 95% LoA ranged from -0.560 to 0.294 [Table 2 and Figure 3]. These relatively wide LoA suggest that individual patients with similar dialysis delivery may demonstrate clinically different Kt/V values depending on the method used, potentially affecting the adequacy classification.
Cross-tabulation revealed that the OCM Kt/V method classified a higher proportion of cases as adequate compared to the Daugirdas Kt/V, causing overestimation (Kappa = 0.45, p <0.0002) [Table 3]. The greatest number of discordant cases was observed between OCM Kt/V and the Daugirdas formula, with 12 patients classified differently by the two methods. OCM Kt/V and nomogram Kt/V showed 9 discordant cases, while the nomogram and Daugirdas formula demonstrated the closest agreement, with only 3 discordant cases. These findings indicate that OCM Kt/V and Daugirdas Kt/V are not fully interchangeable for assessing dialysis adequacy.
| Methods | Adequate (n=50) | Proportion (%) |
|---|---|---|
| OCM | 40 | 80 |
| Nomogram | 33 | 66 |
| Daugirdas formula | 30 | 60 |
OCM: Online clearance monitoring
Subgroup analysis revealed bias values of 0.11 in males and 0.183 in females, suggesting greater overestimation in females. The 95% LoA ranged from -0.299–0.518 in males and from -0.254–0.619 in females. Although visual inspection of the plots also suggested a possible trend, linear regression analysis of the difference against the mean did not demonstrate statistically significant proportional bias in either males (β₁ = −0.130, p = 0.448) or females (β₁ = 0.094, p = 0.647).
These findings indicate that the observed differences between methods are not dependent on the magnitude of Kt/V and instead represent a consistent systematic bias [Figure 4].

Discussion
There have been very few studies validating OCM Kt/V by the UV absorbance method.12,14,15 Moreover, there has been only a single study done in India comparing OCM Kt/V by UV absorbance to other blood-based methods.16 We thus undertook this study to determine whether OCM Kt/V by UV absorbance is in agreement with Kt/V measured by other methods. In our study, the mean OCM Kt/V was 1.43, which was significantly higher, thus overestimating Kt/V compared to the other methods, and this finding was consistent with the study done by Anandi et al. in India.16
The well-known Daugirdas formula is widely used as a method of estimating Kt/V. The nomogram is a derived method based on similar principles. These blood-based methods are cumbersome, limiting their use in daily practice and hence compromising the assessment of HD adequacy, especially in developing countries, which have standalone dialysis centres with limited manpower. The OCM method avoids the limitations of multiple blood samplings and tedious calculations and was developed as a non-invasive and real-time session-by-session method of estimating Kt/V.15 The first commercially available and widely used method of estimating Kt/V by OCM was based on ionic dialysance developed by Fresenius Medical Care, done by measuring the difference in the conductivity of dialysate entering and leaving the dialyser after intermittently altering the conductivity, typically every 45 min.17,18 This method is based on the principle that ionic dialysance of sodium and the effective urea clearance are equivalent. A number of studies have also confirmed the comparable diffusion characteristics of sodium and urea across dialysis membranes and hence their equivalence.19
As far as the ionic dialysance-based method is concerned, there have been enough published studies comparing OCM Kt/V by ionic dialysance to blood-based methods.20-24 Study by Fazl-E-Mateen et al.20 in 2022 compared Kt/V by the Daugirdas formula, nomogram, URR, and OCM by ionic dialysance and showed that OCM slightly underestimated Kt/V as compared to the Daugirdas formula and nomogram. Similarly, other studies comparing Kt/V by ionic dialysance to formula method, such as by Alayoud et al. (Kt/V–1.56 vs. 1.8),21 Baloğlu et al. (Kt/v- 1.37 vs. 1.54),22 Begum et al. (Kt/V-1.2 vs. 1.4),23 have concluded that Kt/V by ionic dialysance-based method underestimates Kt/V. These studies have largely attributed this underestimation to errors in calculating V, which uses various anthropometric formulas such as the Watson formula.11 Lindley et al. have shown that V assessed by the Watson formula leads to underestimation of Kt/V.7
As V tends to be lower in females for any given age, height, and weight because of less muscle mass and less TBW, consequently, less volume of distribution for soluble uremic toxins and less fat mass, thus a smaller ability to adsorb uremic toxins. Thus, the differences in body composition and metabolic rate may have an influence on the production of uremic toxins.25 It signifies that resting energy metabolism, fat-free mass, and uremic toxin production rate might be higher in patients with low body mass.26,27 Therefore, there is a need for higher dialysis dosing in patients with ESRD with higher resting metabolic rate. Spalding et al. found that Kt/V underestimates HD dose in women and small men and have emphasized the need for changing V as a normalising factor to a parameter reflecting metabolic rate, such as body surface area.28
The OCM method by UV absorbance is devoid of the shortcomings of the other methods. This method was first demonstrated by Gál et al. in 1983.29 Castellarnau et al., quoted that UV absorbance adequately represents urea concentration and was less sensitive to measurement errors.15 This method offers the advantage of eliminating the need for V and directly obtaining Kt/V from the decay absorbance curve.
In our study, agreement analysis identified systematic bias, which is further supported by cross-tabulation analysis, which demonstrated that OCM classified a higher proportion of patients as achieving adequate dialysis compared to Daugirdas Kt/V. Hence, the observed overestimation by OCM Kt/V is interpreted as method-related variations rather than any definitive differences in delivered dialysis dose.
In our study, linear regression analysis demonstrated higher Kt/V in females and males with lower body weight. Although visual inspection of the plots also suggested a possible trend, linear regression analysis of the difference against the mean did not demonstrate statistically significant proportional bias. However, proportional bias could not be ruled out definitively due to the small sample size.
Therefore, our observation is that OCM Kt/V and Daugirdas Kt/V methods are not directly interchangeable. The presence of systematic bias, even in the absence of proportional bias, suggests that relying on the OCM method alone may lead to overestimation of dialysis adequacy.
Our study had the following limitations: This was a single-centre study with a small sample size (n = 50). The sample size estimation was based on expected differences in Kt/V values reported in previous studies12,15,16 and was not specifically based on agreement analysis, limiting the precision of agreement estimates. Although regression analysis did not demonstrate statistically significant proportional bias, the relatively small sample size may have limited the statistical power to detect subtle proportional trends. In the absence of an external reference standard, this study could not determine which method most accurately reflects true dialysis adequacy. In this study, only the B Braun dialog HD machine was used with highly standardized parameters such as fixed blood flow rate, dialysate flow rate, duration of HD session, and single dialyzer surface area; their effect on Kt/V couldn’t be studied, and also prevented external generalizability to other platforms and more heterogeneous HD prescriptions. In conclusion, we found that OCM Kt/V by UV absorbance overestimates Kt/V compared to Daugirdas formula and nomogram Kt/V, with moderate agreement, limiting its interchangeability.
Author contributions
Conceptualization, research framework development, and supervision: AJ; Data validation, literature review, and manuscript drafting: MTP, SR, SBR, SVA; Project administration, regulatory compliance, manuscript review and editing: AJ, PK; Methodology development: MTP, AB, SD; Patient recruitment, database management: SD; Study design: MTP; Data analysis: MTP, SD; Critical revision: MTP; Data collection: AB, SD, MTP, SR, SBR, SVA; Interpretation of results: AJ, MTP. All authors participated in data interpretation, critically reviewed the manuscript for important intellectual content, approved the final version for publication, and agreed to be accountable for all aspects of the work.
Conflicts of interest
There are no conflicts of interest.
The authors declare that no generative AI or AI-assisted tools were used in drafting, editing, or preparing this manuscript.
References
- A mechanistic analysis of the national cooperative dialysis study (NCDS) Kidney Int. 1985;28:526-34.
- [CrossRef] [PubMed] [Google Scholar]
- Kt/V (and especially its modifications) remains a useful measure of hemodialysis dose. Kidney Int. 2015;88:466-73.
- [CrossRef] [PubMed] [Google Scholar]
- Quantification of hemodialysis dose: What Kt/V to choose? Int J Artif Organs. 2014;37:29-38.
- [CrossRef] [PubMed] [Google Scholar]
- Routine online assessment of dialysis dose: Ionic dialysance or UV-absorbance monitoring? Semin Dial. 2021;34:116-22.
- [CrossRef] [PubMed] [Google Scholar]
- Dialysis dose (Kt/V) and clearance variation sensitivity using ultraviolet absorbance, blood urea, dialysate urea and ionic dialysance. Nephrol Dial Transplant. 2006;21:2225-31.
- [CrossRef] [PubMed] [Google Scholar]
- Ionic dialysance and assessment of Kt/V: Influence of different estimates of V. Nephrol Dial Transplant. 2007;22:2276-82.
- [CrossRef] [PubMed] [Google Scholar]
- Comparison of methods for determining urea distribution volume. Nephrol Dial Transplant. 2009;24:211-16.
- [CrossRef] [PubMed] [Google Scholar]
- Hemodynamic instability during dialysis: The potential role of intradialytic exercise. Biomed Res Int. 2018;2018:8276912.
- [CrossRef] [PubMed] [PubMed Central] [Google Scholar]
- Volume balance and intradialytic ultrafiltration rate in the hemodialysis patient. Curr Heart Fail Rep. 2017;14:421-7.
- [CrossRef] [PubMed] [PubMed Central] [Google Scholar]
- Improved equation for estimating single-pool Kt/V. Nephrol Dial Transplant. 2013;28:2156-60.
- [CrossRef] [PubMed] [PubMed Central] [Google Scholar]
- A nomogram approach to hemodialysis urea modeling. Am J Kidney Dis. 1994;23:33-40.
- [CrossRef] [PubMed] [Google Scholar]
- Estimation of dialysis dose by UV absorbance. Am J Kidney Dis. 2003;41:1026-36.
- [CrossRef] [PubMed] [Google Scholar]
- Total body water volumes estimated from anthropometric measurements. Am J Clin Nutr. 1980;33:27-39.
- [CrossRef] [PubMed] [Google Scholar]
- Dialysis adequacy. In: Applied home hemodialysis Applied home hemodialysis. Springer nature Switzerland; 2025. p. :33-4.
- [Google Scholar]
- Real-time Kt/V determination by UV absorbance. Kidney Int. 2010;78:920-25.
- [CrossRef] [PubMed] [Google Scholar]
- Assessment of hemodialysis adequacy by online clearance monitoring. G Ital Nefrol. 2024;41:2024-vl6.
- [CrossRef] [PubMed] [Google Scholar]
- Conductivity monitoring in haemodialysis sessions. Nephrol Dial Transplant. 1999;14:2607-2613.
- [CrossRef] [PubMed] [Google Scholar]
- Relationship between ionic dialysance and urea clearance. Am J Kidney Dis. 2001;38:565-74.
- [CrossRef] [PubMed] [Google Scholar]
- Assessment of haemodialysis adequacy by ionic dialysance. Nephrol Dial Transplant. 2003;18:559-62.
- [CrossRef] [PubMed] [Google Scholar]
- Comparison between different methods of calculating Kt/V. Pak J Med Sci. 2022;38:167-71.
- [CrossRef] [PubMed] [PubMed Central] [Google Scholar]
- Kt/V by ionic dialysance: Interpretation limits. Indian J Nephrol. 2012;22:333-39.
- [CrossRef] [PubMed] [PubMed Central] [Google Scholar]
- Begum M, Arshad AR. Comparison of Kt/V estimation methods 2020.
- Online conductivity monitoring of dialysis adequacy. Int J Nephrol Renovasc Dis. 2009;2:27-31.
- [CrossRef] [PubMed] [Google Scholar]
- Determination of urea distribution volume. Kidney Int. 2003;64:2262-71.
- [CrossRef] [PubMed] [Google Scholar]
- Hemodialysis. In: Comprehensive clinical nephrology Comprehensive clinical nephrology. Elsevier; 2010. p. :1053-9.
- [Google Scholar]
- Comparison of dialysis dose through real-time Kt/V by ultraviolet absorbance of spent dialysate, single-pool Daugirdas II, and Kt/BSA according to sex and age. J Bras Nefrol.. 2021;43:52-60.
- [CrossRef] [PubMed] [PubMed Central] [Google Scholar]
- Continuous monitoring of haemodialysis efficiency. Acta Chir Hung. 1983;24:231-9.
- [PubMed] [Google Scholar]